The Model 8630-SM PresSura Room Pressure Monitor is designed to measure the pressure differential in hospital rooms.
The 8630-SM measures the actual room pressure differential using TSI’s unique, bi-directional pressure sensor. If the pressure differential is too great, too small, or the wrong direction, the Model 8630-SM will activate audible and visual alarms. The Model 8630-SM features adjustable alarm delays to avoid nuisance alarms. When the room is unoccupied, the Model 8630-SM can be changed to no isolation mode, de-activating all alarms. Hie Model 8630-SM easily integrates to the building management sys¬tem using a low alarm relay and an analog output of the measured pressure differential.
Display Range -0.20000 to +0.20000 in. H20
Low Alarm Range -0.19500 to +0.19500 in. H2O
High Alarm Range -0.19500 to +0.19500 in. H20
Low Alarm Contact SPST (NO) Max current 5A, max voltage 150VDC,
250VAC Min switch load 10 mA, 5VDCC
User Configurable Analog Output Signal: 0 to 10V or 4 to 20 mA Range: -0.100 to +0.100 in.
H2O or -0.0100 to +0.0100 in. H20
Operating Temperature 32 to 120T
Input Power Weight 24VAC, 5 W max 0.7 lb
Range -0.20000 to +0.20000 in. H2O
Accuracy ±10% of reading
±0.00001 in. H2O
Resolution 55 to 95T
Temp. Comp. Range 0.16 W at 0 in. H2O 0.20 W at 0.00100 in. H2O
Power 0.2 lb